Repair record
Fault
25 error message report on our systems, no scan image at all. Abnormality exists in the angle control
Repair Advice
Replace the crystal, repair circuit, repair the angle problem.
Measurements
| Cable | damaged outer skin;; Yellowing |
| Circuit | Unable to judge |
| Probe casing | |
| Strain relief | OK |
| Matching Layer |
Verify
Scan to verify
https://www.medtrade.world/verification/geprobe/philips-x8-2t-repair-case-21
Buyers scan the QR or paste the URL into medtrade.world/v and see the canonical report — matching the printed PDF byte-for-byte.
Photos
Fault photos
Post-repair photos
Pre-repair measurements
| Cable | damaged outer skin;; Yellowing |
|---|---|
| Circuit | Unable to judge |
| Probe casing | |
| Strain relief | OK |
| Matching Layer | |
| Connector casing | slightly cracked |
| Crystal detection | fault |
| Scan image result | fault |
| Can test on system | No |
| Lens/Membrane/Dome | Scratches |
| Other descriptions | |
| Had ever been repaired by others? | Yes, but not fixed. |